UPC models

Description

For comparative measurements

 

Specification

Standard EN ISO 3650
Maximum permissible errors Provided all the metrological conditions are met, the reliability of the comparator during direct steel gauge block measurement is expressed as follows: WITHOUT HIGH PRECISION OPTION (without calibration certificate) - Repeatability limit (with no effect due to external temperature): 0,025 μm - Measurement uncertainty: +/- (0,10 + 1,0 x L) µm (L in m) Condition involves the use of reference standards whose uncertainty is as follows: - U ≤ ± 0,030 μm when calibrating the comparator - U ≤ ± (0,05 + 0,5 · L) μm (L in m) when calibrating the gauge blocks WITHOUT HIGH PRECISION OPTION - Repeatability limit (with no effect due to external temperature): 0,015 μm - Measurement uncertainty: +/- (0,05 + 0,5 x L) µm (L in m) Condition involves the use of reference standards whose uncertainty is as follows: - U ≤ ± 0,015 μm when calibrating the comparator - U ≤ ± (0,02 + 0,2 · L) μm (L in m) when calibrating the gauge blocks
Weight UPC: 23 kg UPT (temperature module): 4 kg
Function(s) Measuring configuration Two probes with mechanical contact with the measuring face to be probed are connected in sum measurement (function +A+B). Measuring points On the reference gauge block: at the centre of the measuring face (point R). On the gauge block to be measured: at the centre (point 1) as well as the four corners of the measuring face, each lying 2 mm away from the adjacent faces (points 2 to 5). The central length lc is determined by probing both points R and 1. For establishing lengths "at any point", the measurements shall be carried out at points R plus 1 to 5. The variation in length v is obtained from measurements taken at points 1 to 5.
Included in delivery On demand: TESA calibration certificate (SCS certificate for the temperature device)